Oil Free Compressor For Oxygen Generator ZW-27/1.4-A

Short Description:


Product Detail

Product Tags

Product Introduction

Product Introduction

①. Basic parameters and performance indicators
1. Rated voltage/frequency:AC 220V/50Hz
2. Rated current:0.7A
3. Rated power:150W
4. Motor stage:4P
5. Rated speed:1400RPM
6. Rated flow:≥27L/min
7. Rated pressure:0.14MPa
8. Noise:<59.5dB(A)
9. Operating ambient temperature:5-40℃
10. weight:2.8KG
②. Electrical performance 
1. Motor temperature protection:135℃
2. Insulation class:class B
3. Insulation resistance:≥50MΩ
4. Electrical strength :1500v/min( No breakdown and flashover )
③. Accessories 
1. Lead length : Power-line length 580±20mm, Capacitance-line length 580+20mm
2. capacitance :450V 3.55µF
3. Elbow:G1/8
④. Test method
1. Low voltage test :AC 187V. Start the compressor for loading, and do not stop before the pressure rises to 0.1MPa
2. Flow test : Under the rated voltage and 0.14MPa pressure, start to work to a stable state, and the flow reaches 27L/min.

Product Indicators

Model

Rated voltage and frequency

Rated power(W)

Rated current(A)

Rated working pressure(KPa)

Rated volume flow

(LPM)

capacitance(μF)

noise(㏈(A))

Low pressure start(V)

Installation dimension(mm)

Product dimensions(mm)

weight(KG)

ZW-27/1.4-A

AC 220V/50Hz

150W

0.7A

1.4

≥27L/min

4.5μF

≤48

187V

102×73

153×95×136

2.8

Product Appearance Dimensions drawing: (Length: 153mm × Width: 95mm × Height: 136mm)

img-1

Oil-free compressor( ZW-27/1.4-A )for oxygen concentrator

1. Imported bearings and sealing rings for good performance.
2. Less noise, suitable for long-term operation.
3. Applied in many fields.
4. Durable.

 

Compressor common fault analysis
1. Insufficient exhaust volume
Insufficient displacement is one of the most prone failures of compressors, and its occurrence is mainly caused by the following reasons:
1. The fault of the intake filter: fouling and clogging, which reduces the exhaust volume; the suction pipe is too long and the pipe diameter is too small, which increases the suction resistance and affects the air volume, so the filter should be cleaned regularly.
2. The reduction of the compressor speed reduces the displacement: the air compressor is used improperly, because the displacement of the air compressor is designed according to a certain altitude, suction temperature and humidity, when it is used on a plateau exceeding the above standards When the suction pressure decreases, the displacement will inevitably decrease.
3. The cylinder, piston, and piston ring are severely worn and out of tolerance, which increases the relevant clearance and leakage, which affects the displacement. When it is normal wear and tear, it is necessary to replace the wearing parts in time, such as piston rings. It belongs to incorrect installation, if the gap is not suitable, it should be corrected according to the drawing. If there is no drawing, experience data can be taken. For the gap between the piston and the cylinder along the circumference, if it is a cast iron piston, the gap value is the diameter of the cylinder. 0.06/100~0.09/100; for aluminum alloy pistons, the gap is 0.12/100~0.18/100 of the diameter of the gas diameter; steel pistons can take the smaller value of aluminum alloy pistons.